3310A

PRECISION Multiproduct Calibrator

  • 0 – 1000V DC / AC Voltage
  • 0 – 30A DC / AC Current
  • 0 – 1 GOhm Resistance
  • 1nF – 10mF Capacitance
  • 1Hz – 10MHz Frequency
  • Range of external accessories to increase workload

ASK AN EXPERT

Information

Description

The 33XXA Series multiproduct calibrators are modular versions of the class leading 3000A Series Calibrators. Designed to reduce cost, the 33XXA Series feature 0 – 20V DC output as standard with a wide range of options including

  • 20V – 1000V High Voltage Output
  • 0mA – 200mA Low Current Output
  • 200mA – 30A (3310A/3341A) High Current Output
  • AC Output (includes all previously specified options)
  • Resistance
  • Capacitance

The configurability suits specialised applications where only limited functions are required, for example ATE systems or Dual Source resistance measurements systems as used by NMI’s

3310A Specifications

3310A Specifications

DC VOLTAGE

Best 1 Year Total Accuracy

0 - ±20V (0 - ±1000V Option)
± 8 ppm

DC CURRENT

Best 1 Year Total Accuracy

0 - ±30A
±50ppm

AC VOLTAGE

Best 1 Year Total Accuracy

20mV - 1000V, 10Hz - 1MHz
±0.015%

AC CURRENT

Best 1 Year Total Accuracy

20uA - 30A, 10Hz - 10kHz
0.04%

RESISTANCE (PASSIVE) (OPTIONAL)

Best 1 Year Total Accuracy

0 - 1 GOhms
18ppm

CAPACITANCE (PASSIVE) (OPTIONAL)

Best 1 Year Total Accuracy

1nF - 10uF
0.25%

3000A Calibration Resources

3000A Calibration Resources

Transmille Product Guide
Transmille Warranty

Transmille Warranty

Options & Accessories

External

3000 Lead

A comprehensive set of test leads providing every connection required for typical measurements

External

3xxx Soft Case

A lightweight solution for transporting your investment

Designed for transporting your calibrator through couriers and harsh environments, with custom laser cut foam for holding units securely

A superior thermocouple simulation/measurement option, using a unique external adapter.

  • Offers fast and accurate calibration of thermocouple meters
  • Uses an innovative external measurement of cold junction temperature for ultimate accuracy
  • Utilising a thermocouple plug removes the requirement for expensive thermocouple compensation cable
 

Calibrate clamp meters at up to 1500A using the EA002 clamp meter. A unique design enables clamps of all sizes to be used, while minimising errors due to alignment

  • Calibrates clamp meters up to 1500A (30A input)
  • 2/10/50 Turn Coils
  • High accuracy AC/DC design, with minimal errors due to alignment
  • Wide range of clamp sizes covered, from 10mm jaw diameter to larger clamps
  • Alignment table included for accurate and repeatable readings

Instantly add the ability to calibrate tachometers to 60,000 RPM by adding the EA003 to your calibrator

  • High-intensity LED light source
  • Direct keyboard input in RPM
  • Control via ProCal for fully automated calibration
  • Output range from 60 to 60,000 RPM

Adds measurement capability to 1pA resolution, and measures high value resistors to 1TOhm using the calibrator as a precision DC voltage source

  • Accurate current measurement down to 1pA resolution
  • Ideal for high value resistance measurements up to 1 TOhm
  • High resistance measurement voltages up to 1000V
  • 5 current measurement rnages from 10nA to 100uA
  • Control directly from the calibrator front panel or use with ProCal software

As multimeters add more accurate low current ranges, traditional multi-product calibrators lack the precision and noise level for accurate calibration. With the EA013 pA source adapter, calibration of these instruments becomes simple

  • Ideal for calibration of electrometer instruments
  • Source accurate current down to picoamp levels
  • Low open circuit compliance voltage
  • Safe to use on sensitive input circuits
  • Controlled directly from the calibrator front panel or via supplied software
  • Compatible with ProCal for full measurement automation

Extend your laboratory's capabilities with the ability to calibrate torque screwdrivers and much more

Combining Transmille's most popular options, the EA015 provides a cost-effective method of increasing the workload coverage of Transmille calibrators

  • Insulation Testers
  • Continuity Testers
  • Optical Tachometers
  • Thermocouple Meters
  • Clamp Meters
  • Source functions of process control meters

Measure environmental factors both on site and in the laboratory

  • Humidity measurement from 10% to 90%
  • Temperature measurement from 0°C to 50°C

Automate calibration of 4 channel oscilloscopes with automised switching of channels with the EA017 adapter

  • 4 Output channels for automating oscilloscope calibration
  • Minimises lead changing when calibrating oscilloscopes
  • Choose between direct or buffered outputs for amplitude calibration