4000 Series

Advanced Multiproduct Calibrator

The 4000 Series is aimed at laboratories, manufacturers and design facilities requiring a wide range of outputs with accuracies up to 8ppm.

The 4000 Series builds on the trusted reliability of the 3000A series, offering an updated user interface accessed through a large 7" capacitive touch screen to enable advanced functionality, including a built-in procedure mode and inbuilt connection prompts to minimise errors.

Procedures can be loaded into the 4000 Series via USB memory stick, with results stored for integration back into ProCal software. The 4000 Series also provides USB printer connectivity as standard, supporting a wide range of PCL compatible printers. Reports can be printed directly from the calibrator - ideal for in-situ or on-site applications.

  • 0 - 1000V DC / AC Voltage
  • 0 - 30A DC / AC Current
  • Variable Resistance up to 1GOhm
  • Variable Capacitance to 100mF
  • 1Hz - 10MHz Frequency
  • AC current maximum frequency 30kHz
  • Range of external accessories to increase workload
  • USB and GPIB interface fitted as standard
  • Removable Handles

ASK AN EXPERT

Highlights Product Range Resources Videos & Demos

HIGHLIGHTS

Traceable Self-Verification Functionality

To maintain confidence in your equipment, simply connect an 8000 series digital multimeter to the rear of the unit and enter 'self-calibration mode'. The 4000 series will then proceed to measure its outputs automatically and generate a test report which can be exported to a USB memory stick for analysis on a computer, and print a test report via a USB printer for maintaining records.

SOFTWARE COMPATABILITY

As with all Transmille products, compatibility with ProCal calibration software is guaranteed. Existing procedures will not need to be re-written due to Transmille's unique implementation of flexible standards support where procedures are written around the UUT not the standard.

PRODUCT RANGE

4010

4015

DC VOLTAGE

Best 1 Year Total Accuracy

0 - ±1025V
±8ppm

0 - ±1025V
±15ppm

DC CURRENT

Best 1 Year Total Accuracy

0 - 30A
±50ppm

0 - 30A
±50ppm

AC VOLTAGE

Best 1 Year Total Accuracy

20mV - 1025V, 10Hz - 1MHz
±0.015%

20mV - 1000V, 10Hz - 500kHz
±0.015%

AC CURRENT

Best 1 Year Total Accuracy

20uA - 30A, 10Hz - 30kHz
0.04%

20uA - 30A, 10Hz - 30kHz
0.04%

RESISTANCE (PASSIVE)

Best 1 Year Total Accuracy

0 - 1 GOhms
18ppm

0 - 1 GOhms
40ppm

FREQUENCY

Best 1 Year Total Accuracy

1Hz - 10MHz
1ppm

1Hz - 10MHz
20ppm (1ppm Option)

TEMPERATURE (THERMOCOUPLE)

Best 1 Year Total Accuracy

J / K / T / R / S / B / U / N / E / L / U / C / 10uV°C
±0.09°C

J / K / T / R / S / B / U / N / E / L / U / C / 10uV°C
±0.09°C

RESISTANCE (SIMULATED)

Best 1 Year Total Accuracy

0 Ohms - 1 GOhm
100ppm

0 Ohms - 1 GOhm
100ppm

4010

4015

Find Out More
Find Out More

Resources

Transmille Warranty

VIDEOS & DEMOS

4000 SERIES - DIRECT OHMS SELECTION

In this short video we demonstrate the latest updates to the 4000 series. The addition of new function keys enable direct access to the decade passive resistors enabling outputs to be directly accessed with a single touch of the screen.

4000 Series - Automated Verification

An example of the automated verification functionality using a Transmille 8081 precision multimeter.

4000 Series - Procedure Mode

Demonstration of the built-in procedure mode using a typical multimeter.

Options & Accessories

An external pressure module for use with Transmille products, designed to extend capabilities to 100 Bar

An external pressure module for use with Transmille products, designed to extend capabilities from -1 to +1 Bar

An external pressure module for use with Transmille products, designed to extend capabilities from -1 to +1 Bar

Add RS232 support to modern computers and laptops